Each time you perform a zero calibration, the instrument stores a new set of offset correction constants for every measurement function and range. The instrument sequences through all required functions and ranges automatically and stores new zero offset calibration constants. All offset corrections are determined automatically. You may not correct a single range or function without re–entering ALL zero offset correction constants automatically. This feature is intended to save calibration time and improve zero calibration consistency.
![]() |
Never turn off the instrument during Zero Calibration. This may cause ALL calibration memory to be lost. |
This calibration procedure applies to all models (34460A/61A/65A/70A).
Be sure to allow the instrument to warm up and stabilize for 90 minutes before performing the calibration.
![]() |
For the 34461A/65A/70A, two shorting blocks are recommended; one on the front panel and one on the rear panel. A single shorting block can be used with the 34460A. |
![]() |
The following gain and flatness calibration procedures can be done using either the front or rear terminals (rear terminals not available on the 34460A). On the 34461A/65A/70A, make sure the Front/Rear switch matches the terminals in use. |
Next Procedure |
![]() |